Purpose:
The Programme aims to assess the performance of organisations utilising blood gas analysers for the analysis of the patients’ haemoglobin and co-oximetry status.
The Programme is available to both Laboratory and Point of Care testing sites and is part of the Blood Gas and Co-oximetry Programme offering the complete repertoire of blood gas analytes including calculated parameters, electrolytes, metabolites and Co-oximetry.
Scope:
The material is a blood-based matrix containing the 4 major haemoglobin derivates as well as Bilirubin and unlike dyes provided by other EQA providers has equivalent spectral properties to whole blood and is therefore compatible with all major Co-oximeters. Three samples are distributed monthly, with a minimum of 33 samples distributed over the year covering a wide clinically relevant range. The samples are packaged in an easy to use dual-chambered vial that separates the oxyhaemoglobin from methaemoglobin to enhance stability.
The samples consist of a panel of 6 linearly related samples distributed on a number of occasions over that period which are used to assess both laboratory and method performance, including linearity, bias, within and between batch imprecision. Once a year, tonometered whole blood is also distributed which allows for the assessment of the derived parameter Oxygen Saturation. Participants can also select to receive 3 samples, 6 times a year.
![]() |
Key Features:
|
| Co-oximetry Programme | ||
|---|---|---|
| Analyte | Approx. Range Covered | |
| Total Haemoglobin (ctHB) | 100 - 190 | g/L |
| Carboxyhaemoglobin (FCOHb) | 1 - 32 | % |
| Methaemoglobin (FMetHb) | 2 - 35 | % |
| Oxyhaemoglobin (FO2Hb) | 35 - 95 | % |
| Oxygen saturation (sO2) Annual assessment | 90 - 100 | % |
| Bilirubin (ctBil) | 0 - 100 | µmol/L |
Related Resources
Last updated: 18/09/2025